AIM:To evaluate the accuracy of axial length(AL)measurements obtained from immersion B-scan ultrasonography(immersion B-scan)for intraocular lens(IOL)power calculation in patients with high myopia and cataracts.METHODS:Immersion B-scan,contact A-scan ultrasonography(contact A-scan),and the IOLMaster were used to preoperatively measure the AL in 102 eyes from 102 patients who underwent phacoemulsification and IOL implantation.Patients were divided into two groups according to the AL:one containing patients with22 mm≤AL<26 mm(group A)and the other containing patients with AL≥26 mm(group B).The mean error(ME)was calculated from the difference between the AL measurement methods predicted refractive error and the actual postoperative refractive error.RESULTS:Ingroup A,ALs measured byimmersion Bscan(23.48±1.15)didn’t differ significantly from those measured by the IOLMaster(23.52±1.17)or from those by contact A-scan(23.38±1.20).In the same group,the standard deviation(SD)of the mean error(ME)of immersion B-scan(-0.090±0.397 D)didn’t differ significantly from those of IOLMaster(-0.095±0.411 D)and contact A-scan(-0.099±0.425 D).In group B,ALs measured by immersion B-scan(27.97±2.21 mm)didn’t differ significantly from those of the IOLMaster(27.86±2.18 mm),but longer than those measured by Contact A-scan(27.75±2.23 mm,P=0.009).In the same group,the standard deviation(SD)of the mean error(ME)of immersion B-scan(-0.635±0.157 D)didn’t differ significantly from those of the IOLMaster(-0.679±0.359 D),but differed significantly from those of contact A-scan(-0.953±1.713 D,P=0.028).CONCLUSION:ImmersionB-scanexhibitsmeasurement accuracy comparable to that of the IOLMaster,and is thus a good alternative in measuring AL in eyes with high myopia when the IOLMaster can’t be used,and it is more accurate than the contact A-scan.
·AIM: To evaluate the accuracy of 20 MHz immersion Bscan ultrasonography in observing lens and to investigate the value of this noninvasive preoperative diagnosis method in alkali burn eyes.·METHODS: It was a comparative study. Fifty-six cases(56 eyes) of alkali burn eyes were examined by ultrasound biomicroscopy(UBM) and immersion 20 MHz B-scan ultrasonography from June 2011 to April 2013,the images were analyzed, and the ultrasonographic diagnosis compared with the operation results.·RESULTS: In 56 alkali burn eyes examined by UBM, the lens were not detected in 16 eyes; the IOL could be detected in 2 eyes; the anterior lens capsule surface or/and the front lens could be detected in 18 eyes, and lens opacification in 3 eyes of them; suspected abnormal lens were detected in the other 20 eyes. In all the same eyes examined by immersion 20 MHz B-scan ultrasonography,the lens were not detected in 16 eyes; the IOL could be detected in 2 eyes; 24 abnormal lens(opacity, lens expansion, shrinkage) and 14 normal lens were found.Compared with the intraoperative findings, the diagnostic accordance rate of the immersion 20 MHz B-scan appearance of lens was 100%(56/56), which was significantly higher than examined by UBM 57.14%(32/56)(χ2=30.55, P =0.0000).·CONCLUSION: Immersion 20 MHz B-scan ultrasonography can observe the lens accurately in alkali burn eyes. It has important clinical value to combine with UBM in eyes of alkali burn.